OBJECTIVES : The aim of this study was to evaluate the usefulness of breath-hold turbo spin echo with deep learning-based reconstruction (BH-DL-TSE) in acquiring fat-suppressed T2-weighted images (FS-T2WI) of the liver by comparing this method with conventional free-breathing turbo spin echo (FB-TSE) and breath-hold half Fourier single-shot turbo spin echo with deep learning-based reconstruction (BH-DL-HASTE).

MATERIALS AND METHODS : The study cohort comprised 111 patients with suspected liver disease who underwent 3 T magnetic resonance imaging. Fifty-eight focal solid liver lesions ≥10 mm were also evaluated. Three sets of FS-T2WI were acquired using FB-TSE, prototypical BH-DL-TSE, and prototypical BH-DL-HASTE, respectively. In the qualitative analysis, 2 radiologists evaluated the image quality using a 5-point scale. In the quantitative analysis, we calculated the lesion-to-liver signal intensity ratio (LEL-SIR). Friedman test and Dunn multiple comparison test were performed to assess differences among 3 types of FS-T2WI with respect to image quality and LEL-SIR.

RESULTS : The mean acquisition time was 4 minutes and 43 seconds ± 1 minute and 21 seconds (95% confidence interval, 4 minutes and 28 seconds to 4 minutes and 58 seconds) for FB-TSE, 40 seconds for BH-DL-TSE, and 20 seconds for BH-DL-HASTE. In the qualitative analysis, BH-DL-HASTE resulted in the fewest respiratory motion artifacts (P < 0.0001). BH-DL-TSE and FB-TSE exhibited significantly less motion-related signal loss and clearer intrahepatic vessels than BH-DL-HASTE (P < 0.0001). Regarding the edge sharpness of the left lobe, BH-DL-HASTE scored the highest (P < 0.0001), and BH-DL-TSE scored higher than FB-TSE (P = 0.0290). There were no significant differences among 3 types of FS-T2WI with respect to the edge sharpness of the right lobe (P = 0.1290), lesion conspicuity (P = 0.5292), and LEL-SIR (P = 0.6026).

CONCLUSIONS : BH-DL-TSE provides a shorter acquisition time and comparable or better image quality than FB-TSE, and could replace FB-TSE in acquiring FS-T2WI of the liver. BH-DL-TSE and BH-DL-HASTE have their own advantages and may be used complementarily.

Ichinohe Fumihito, Oyama Kazuki, Yamada Akira, Hayashihara Hayato, Adachi Yasuo, Kitoh Yoshihiro, Kanki Yusuke, Maruyama Katsuya, Nickel Marcel Dominik, Fujinaga Yasunari

2022-Dec-27