Receive a weekly summary and discussion of the top papers of the week by leading researchers in the field.

In Magnetic resonance in medicine ; h5-index 66.0

PURPOSE : To substantially shorten the acquisition time required for quantitative three-dimensional (3D) chemical exchange saturation transfer (CEST) and semisolid magnetization transfer (MT) imaging and allow for rapid chemical exchange parameter map reconstruction.

METHODS : Three-dimensional CEST and MT magnetic resonance fingerprinting (MRF) datasets of L-arginine phantoms, whole-brains, and calf muscles from healthy volunteers, cancer patients, and cardiac patients were acquired using 3T clinical scanners at three different sites, using three different scanner models and coils. A saturation transfer-oriented generative adversarial network (GAN-ST) supervised framework was then designed and trained to learn the mapping from a reduced input data space to the quantitative exchange parameter space, while preserving perceptual and quantitative content.

RESULTS : The GAN-ST 3D acquisition time was 42-52 s, 70% shorter than CEST-MRF. The quantitative reconstruction of the entire brain took 0.8 s. An excellent agreement was observed between the ground truth and GAN-based L-arginine concentration and pH values (Pearson's r > 0.95, ICC > 0.88, NRMSE < 3%). GAN-ST images from a brain-tumor subject yielded a semi-solid volume fraction and exchange rate NRMSE of 3 . 8 ± 1 . 3 % $$ 3.8\pm 1.3\% $$ and 4 . 6 ± 1 . 3 % $$ 4.6\pm 1.3\% $$ , respectively, and SSIM of 96 . 3 ± 1 . 6 % $$ 96.3\pm 1.6\% $$ and 95 . 0 ± 2 . 4 % $$ 95.0\pm 2.4\% $$ , respectively. The mapping of the calf-muscle exchange parameters in a cardiac patient, yielded NRMSE < 7% and SSIM > 94% for the semi-solid exchange parameters. In regions with large susceptibility artifacts, GAN-ST has demonstrated improved performance and reduced noise compared to MRF.

CONCLUSION : GAN-ST can substantially reduce the acquisition time for quantitative semi-solid MT/CEST mapping, while retaining performance even when facing pathologies and scanner models that were not available during training.

Weigand-Whittier Jonah, Sedykh Maria, Herz Kai, Coll-Font Jaume, Foster Anna N, Gerstner Elizabeth R, Nguyen Christopher, Zaiss Moritz, Farrar Christian T, Perlman Or

2022-Dec-31

chemical exchange saturation transfer, generative adversarial network, magnetic resonance fingerprinting, magnetization transfer, pH, quantitative imaging