Receive a weekly summary and discussion of the top papers of the week by leading researchers in the field.

In European radiology ; h5-index 62.0

BACKGROUND : Deep learning image reconstructions (DLIR) have been recently introduced as an alternative to filtered back projection (FBP) and iterative reconstruction (IR) algorithms for computed tomography (CT) image reconstruction. The aim of this study was to evaluate the effect of DLIR on image quality and quantification of coronary artery calcium (CAC) in comparison to FBP.

METHODS : One hundred patients were consecutively enrolled. Image quality-associated variables (noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR)) as well as CAC-derived parameters (Agatston score, mass, and volume) were calculated from images reconstructed by using FBP and three different strengths of DLIR (low (DLIR_L), medium (DLIR_M), and high (DLIR_H)). Patients were stratified into 4 risk categories according to the Coronary Artery Calcium - Data and Reporting System (CAC-DRS) classification: 0 Agatston score (very low risk), 1-99 Agatston score (mildly increased risk), Agatston 100-299 (moderately increased risk), and ≥ 300 Agatston score (moderately-to-severely increased risk).

RESULTS : In comparison to standard FBP, increasing strength of DLIR was associated with a significant and progressive decrease of image noise (p < 0.001) alongside a significant and progressive increase of both SNR and CNR (p < 0.001). The use of incremental levels of DLIR was associated with a significant decrease of Agatston CAC score and CAC volume (p < 0.001), while mass score remained unchanged when compared to FBP (p = 0.232). The underestimation of Agatston CAC led to a CAC-DRS misclassification rate of 8%.

CONCLUSION : DLIR systematically underestimates Agatston CAC score. Therefore, DLIR should be used cautiously for cardiovascular risk assessment.

KEY POINTS : • In coronary artery calcium imaging, the implementation of deep learning image reconstructions improves image quality, by decreasing the level of image noise. • Deep learning image reconstructions systematically underestimate Agatston coronary artery calcium score. • Deep learning image reconstructions should be used cautiously in clinical routine to measure Agatston coronary artery calcium score for cardiovascular risk assessment.

Rossi Alexia, Gennari Antonio G, Etter Dominik, Benz Dominik C, Sartoretti Thomas, Giannopoulos Andreas A, Mikail Nidaa, Bengs Susan, Maurer Alexander, Gebhard Catherine, Buechel Ronny R, Kaufmann Philipp A, Fuchs Tobias A, Messerli Michael

2022-Dec-08

Cardiac computed tomography, Cardiovascular risk, Coronary artery calcium, Deep learning, Image reconstruction