Receive a weekly summary and discussion of the top papers of the week by leading researchers in the field.

In Journal of medical imaging (Bellingham, Wash.)

Purpose : Chest X-ray (CXR) use in pre-MRI safety screening, such as for lead-less implanted electronic device (LLIED) recognition, is common. To assist CXR interpretation, we "pre-deployed" an artificial intelligence (AI) model to assess (1) accuracies in LLIED-type (and consequently safety-level) identification, (2) safety implications of LLIED nondetections or misidentifications, (3) infrastructural or workflow requirements, and (4) demands related to model adaptation to real-world conditions.

Approach : A two-tier cascading methodology for LLIED detection/localization and identification on a frontal CXR was applied to evaluate the performance of the original nine-class AI model. With the unexpected early appearance of LLIED types during simulated real-world trialing, retraining of a newer 12-class version preceded retrialing. A zero footprint (ZF) graphical user interface (GUI)/viewer with DICOM-based output was developed for inference-result display and adjudication, supporting end-user engagement and model continuous learning and/or modernization.

Results : During model testing or trialing using both the nine-class and 12-class models, robust detection/localization was consistently 100%, with mAP 0.99 from fivefold cross-validation. Safety-level categorization was high during both testing ( AUC 0.98 and 0.99 , respectively) and trialing (accuracy 98% and 97%, respectively). LLIED-type identifications by the two models during testing (1) were 98.9% and 99.5% overall correct and (2) consistently showed AUC 0.92 (1.00 for 8/9 and 9/12 LLIED-types, respectively). Pre-deployment trialing of both models demonstrated overall type-identification accuracies of 94.5% and 95%, respectively. Of the small number of misidentifications, none involved MRI-stringently conditional or MRI-unsafe types of LLIEDs. Optimized ZF GUI/viewer operations led to greater user-friendliness for radiologist engagement.

Conclusions : Our LLIED-related AI methodology supports (1) 100% detection sensitivity, (2) high identification (including MRI-safety) accuracy, and (3) future model deployment with facilitated inference-result display and adjudication for ongoing model adaptation to future real-world experiences.

White Richard D, Demirer Mutlu, Gupta Vikash, Sebro Ronnie A, Kusumoto Frederick M, Erdal Barbaros Selnur

2022-Sep

MRI safety, artificial intelligence, loop recorders, pacemakers